Products PRODUCTS

Product name:PILATUS3 R CdTe for Laboratory
brand:Dectris
Place of Origin:Switzerland
The PILATUS3 R CdTe X-ray detector series combines the advantages of noise-free single-photon counting with the benefits of direct detection of hard X-rays in large cadmium telluride (CdTe) sensors. Direct detection in a hybrid pixel results in sharper signals and better spatial resolution than scintillator-based detectors. The CdTe X-ray detectors also provides almost 100% efficiency for Mo, Ag, and In radiation. Combined with noise-free single-photon counting, these capabilities permit DECTRIS CdTe X-ray detectors to achieve new levels of sensitivity and accuracy in the laboratory.
	
The PILATUS3 R CdTe X-ray detector series combines the advantages of noise-free single-photon counting with the benefits of direct detection of hard X-rays in large cadmium telluride (CdTe) sensors. Direct detection in a hybrid pixel results in sharper signals and better spatial resolution than scintillator-based detectors. The CdTe X-ray detectors also provides almost 100% efficiency for Mo, Ag, and In radiation. Combined with noise-free single-photon counting, these capabilities permit DECTRIS CdTe X-ray detectors to achieve new levels of sensitivity and accuracy in the laboratory.
Laboratory X-ray sources, particularly for hard X-rays, are much weaker than at the synchrotron and longer exposure times are required. Thanks to the absence of dark current and readout noise, PILATUS3 R detectors outperform all other technologies. Single-photon counting is also free of other sources of noise and inaccuracy intrinsic to charge-integrating CMOS detectors. DECTRIS Instant Retrigger technology in combination with an accurate count rate correction have enabled advanced synchrotron high-energy experiments and are hence fully compatible with all laboratory applications.
The PILATUS3 R CdTe detector series offers high-efficiency direct detection and noise-free single-photon counting — a winning combination for your laboratory.
	
技术参数:
| PILATUS3 R CdTe | 1M | 300K | 300K-W | 100K | 
| 探测器模块数量 | 2 × 5 | 1 × 3 | 3 × 1 | 1 × 1 | 
| 有效面积:宽×高 [mm²] | 168.7 x 179.4 | 83.8 × 106.5 | 253.7 x 33.5 | 83.8 x 33.5 | 
| 像素大小 [μm²] | 172 x 172 | |||
| 总像素数量 | 981 x 1043 | 487 × 619 | 1475 × 195 | 487 x 195 | 
| 间隙宽度, 水平/垂直(像素) *每个模块之间水平间隙增加1个像素 | 7* / 17 | - */ 17 | 7* / - | -* / - | 
| 非灵敏区 [ % ] | 7.8 | 5.7 | 1.1 | 0.2 | 
| 缺陷像素 | < 0.1% | |||
| 最大帧频 [Hz] | 5 | 20 | 20 | 20 | 
| 读出时间 [ms] | 7 | |||
| 点扩散函数 | 1 pixel(FWHM) | |||
| 计数器深度 | 20 bits(1,048,576 counts) | |||
| 功耗 [W] | 165 | 30 | 30 | 30 | 
| 尺寸(WHD)[mm³] | 265 x 286 x 455 | 158 x 193 x 262 | 280 x 62 x 296 | 114 x 69 x 118 (探头) | 
| 重量 [kg] | 25 | 7.5 | 7.0 | 0.9(探头) | 
| 模块冷却 | 水冷 | 水冷 | 水冷 | 水冷 | 
| 外接触发电压 | 5V TTL | 5V TTL | \ | \ | 
		 
 
	
		
	
PILATUS3 R CdTe 1M
 
 
	
		
	
		
	
PILATUS3 R CdTe 300K
 
 
	
		
	
		
	
PILATUS3 R CdTe 300K-W
		 
 
	
PILATUS3 R CdTe 100K
		
	